In this episode of the Oil and Gas Measurement Podcast, Weldon Wright sits down with Brian Lund of Norse Technologies to discuss the importance of proper care, handling, and upkeep of calibration equipment used in measurement operations. The conversation highlights common pitfalls, industry practices, and the role of training and certification in maintaining reliable and accurate calibration tools. Listeners gain a high-level perspective on why calibration equipment deserves the same attention as the systems it is used to verify.
Calibration Equipment – Care and Upkeep Show Notes, Links & Insider Terms:
- Calibration Equipment refers to the precision tools (pressure, temperature, electrical, and dimensional) used to verify and adjust measurement devices to ensure accurate readings in oil and gas operations.
- Custody Transfer Measurement is the measurement of hydrocarbons at transfer points where ownership changes, requiring the highest levels of accuracy and traceability.
- Pressure Calibration Equipment includes devices such as pressure calibrators, pressure modules, and deadweight testers used to verify pressure transmitters and gauges.
- Overpressurization is the application of pressure beyond the rated limits of calibration equipment, identified as the leading cause of equipment failure for both electronic modules and mechanical testers.
- Deadweight Tester (DWT) is a primary pressure standard that generates pressure using calibrated weights and a piston assembly, still widely used and valid for high-accuracy pressure calibration.
- PK Deadweight Tester refers to a specific style of deadweight tester discussed in the episode that includes internal orifices and ceramic components vulnerable to debris and damage.
- Pressure Modules are interchangeable electronic pressure-sensing components used in digital calibrators, particularly sensitive to moisture, debris, and overpressure.
- Differential Pressure (DP) refers to the pressure difference between two points, commonly used in flow measurement and highly sensitive to moisture and contamination during calibration.
- Inches of Water (inH₂O) is a low-pressure measurement unit used in DP applications, especially susceptible to errors caused by condensate or debris in calibration equipment.
- Condensate is liquid (often water) that can accumulate in pressure lines or modules and disrupt accurate low-pressure measurements.
- Processed or Line Gas refers to pipeline gas improperly used as a pressure source during calibration, which can introduce debris, moisture, and contaminants into calibration equipment.
- Nitrogen Source is the preferred clean, dry pressure medium used for pressure calibration, particularly for deadweight testers.
- Debris Contamination describes dirt, sand, or particulates entering calibration devices, often causing plugged orifices, unstable readings, or permanent damage.
- Ceramic Ball (Deadweight Tester Component) is a precision component in some deadweight testers that must remain smooth; chips or damage lead to inaccurate pressure generation.
- Calibration Stabilization Time is the required waiting period after a pressure or temperature change to allow the system to reach equilibrium before recording calibration data.
- Flexible Tubing Runs are long hoses used to connect devices under test to calibration equipment, which can introduce stabilization delays and temperature effects.
- Temperature Effects refer to measurement errors caused by environmental temperature changes, such as sunlight heating long pressure or temperature lines.
- Digital Thermometers are modern temperature calibration tools that have largely replaced mercury thermometers in industry use.
- Temperature Probes are sensing elements attached to thermometers; damage or improper storage was identified as the most common failure mode.
- Thermal Well (Thermowell) is a protective sleeve installed in piping that houses temperature probes; proper probe insertion depth is critical for accurate readings.
- Dry Well / Dry Block Calibrator is a temperature calibration source that provides controlled reference temperatures for one-, two-, or three-point temperature calibrations.
- Reference Thermometer is a calibrated temperature standard used to compare against the device under test during verification or calibration.
- One-Point Verification is a basic temperature or pressure check at a single operating condition to confirm readings are within acceptable limits.
- Two-Point / Three-Point Calibration involves verifying and adjusting measurement devices at multiple points across their operating range to ensure linearity and accuracy.
- Calipers are precision dimensional measurement tools used to inspect orifice plates, tubing, and mechanical components, requiring the same care and certification as other calibration equipment.
- Torque Wrenches are tools used to apply consistent bolt torque, particularly important for pressure-containing connections, and require periodic calibration.
- Multimeters are electronic measurement tools used for voltage, current, and resistance verification, which must be calibrated if used for measurement validation.
- Loop Calibrators are devices used to source and measure 4–20 mA signals, commonly overlooked for recertification despite their role in transmitter verification.
- 4–20 mA is the analog electrical signal standard widely used in industrial instrumentation for transmitting process measurements.
- Annual Certification is the required periodic calibration of test equipment by a qualified lab to confirm accuracy against recognized standards.
- Calibration Laboratory (Cal Lab) is a facility that performs certified calibration services; labs must be qualified and verified for compliance with applicable standards.
- API 2011 is an American Petroleum Institute standard governing calibration practices, accuracy ratios, and compliance requirements for measurement equipment.
- As-Found Data represents the condition and accuracy of calibration equipment upon arrival at the calibration lab before adjustments are made.
- As-Left Data documents the condition and accuracy of calibration equipment after calibration or adjustment has been completed.
- Electronic Drift refers to gradual changes in electronic measurement accuracy over time, reinforcing the need for equipment to return from calibration well within tolerance.
- Accuracy Ratio describes the relationship between calibration equipment accuracy and the accuracy of the device under test, a key requirement for API 2011 compliance.
- Stabilization and Equilibrium emphasize the need to allow pressure and temperature systems to normalize before recording calibration data to ensure valid results.
- Equipment Storage and Handling is repeatedly highlighted as the most critical factor in maintaining calibration equipment accuracy, longevity, and reliability.
Calibration Equipment – Care and Upkeep Full Episode Transcript:
Weldon Wright:
Hello and welcome to another episode of the “Oil and Gas Measurement Podcast.”
Sponsored by GCI, the Gas Certification Institute, who have been providing measurement training, measurement SOPs, consulting, and now the Muddy Boots field operation software for the oil and gas industry for over 20 years.
[music]
Announcer:
Welcome to the Oil and Gas Measurement Podcast, where measurement professionals, bubba geeks, and gurus share their knowledge, experience, and likely a tall tale or two on measurement topics for the oil and gas industry.
Now your host, Weldon Wright.
Weldon:
Hello, and welcome to another episode of the Oil and Gas Measurement Podcast. I’m here with Brian Lund today of Norse Technologies. We’re going to be talking about calibration equipment, the care and upkeep of your calibration equipment.
Before we do that, hello, Brian. Tell us a little bit about yourself.
Brian Lund:
Good morning, Weldon.
Yes, I have 30 years of custody transfer measurement calibration equipment sales, in addition to industrial instrumentation and measurement instrumentation for process variable and controls.
I’m currently employed at Norse Technologies, and have been for the last 25 years, since I started Norse in 1997. I started Norse Technologies with the Heise calibrator product line, selling pressure calibrators and the Heist pressure volume controller into the hydrocarbon custody transfer measurement world.
In 1999, I added Ralston Instruments for the management and connection to equipment being calibrated to facilitate that for the users. We added adapters, hoses, and pressure pumps.
From that, we expanded into flow meters, control valves, pressure and temperature transmitters, and here we are today.
Weldon:
Brian, I guess I’m going to claim I’ve known you since probably 2000, when I first called you up and said, “Hey, I need some of these Heise calibrators.” That was probably, yeah, it had to be 2000, probably, so 22 years or so.
Interestingly enough, I know you have a bachelor’s in industrial engineering and a master’s in industrial engineering. I don’t know about the quality of the master’s, but I know that BS comes from A&M.
You and I were both rattling around in industrial engineering at the same time at A&M back in 80-ish. Right?
Brian: Correct.
Weldon:
Yeah, you’d think you’d run into somebody in a place like that, but there’s a couple of students there.
Any rate, what we want to talk about today is, as I mentioned earlier, we’ve known Brian and started buying test equipment from him 22 years ago. I’ve asked him to come on the podcast and talk a little bit about the care and upkeep of our calibration equipment, or the care and feeding of calibration equipment, as I like to say.
For years, our calibration equipment has been one of the most expensive piece of tools in the measurement tech’s truck. Right? But they haven’t always been cared for as well as they should. There’s problems that arise.
Brian, let’s dig into this, and let’s talk about some of the most common problems you see, like with pressure calibration equipment.
Brian:
With pressure calibration equipment, Weldon, the number one problem we see is overpressurization on the electrical side of the modules. It can also happen on the mechanical side with the PK deadweight testers. Overpressurization is the number one failure that we see.
Then there’s some nuisances and some disturbances that occur on the low inches of water, and specifically in the pressure modules measurement. It doesn’t matter whose piece of calibration equipment it is.
If you get water or any condensate that can build up and drop out in these modules, it can actually perturb the inches of water measurement. It can shift it a little bit. You want to make sure that that’s clean and dry, if you see anything variant moving or you can’t get anything to settle and stabilize.
Debris. Sometimes the gas can back vent. The worst thing you could do is use processed or line gas for your pressure cut, for the media, for your pressure calibration. It still occurs. Even though we’ve tried to teach that out and train that out, it still does occur.
Weldon:
I’m surprised by that part, Brian. We’ve been harping on that for years and years.
I know the issue you mentioned with water in our calibration equipment, especially our low end DP equipment, that’s also a battle that we fought and fought and fought. I don’t know how we get that across to some of the folks in the field.
Brian:
It’s probably just training or hands-on training. I know that I’ve been invited in before to companies that utilize our equipment to do initial kickoff training, but also to come back and do follow-up training after a period of time because people/employees change during that time.
A lot of people will stay there for a good amount of time, but there’s always an influx of new people, people transferring in, young people starting out in the industry. After a period of time, I’ve been invited to come in.
Weldon:
Talking about training, of course, GCI does a lot of fundamentals training with measurement, both gas and liquids, there’s some SCADA training also. One thing that I’ve seen is that the training on individual pieces of calibration equipment can vary greatly with the manufacturer.
Overall, I think we as an industry, I don’t know that we do enough training on our calibration equipment ourself. We focus on the procedures to be used, we focus on the hardware we’re installing out in the field, but many times, the calibration equipment, especially the pressure calibration equipment is, the guy gets handed a box and says, here’s your new calibrator, take care of it.
Sometimes that’s about all the training we have. I’ve always been a proponent of what you talk of bringing the manufacturer’s rep in or the manufacturer themselves, and get them to train those technicians, but it’s a hard battle to fight.
Brian:
I actually did that for you several times. It was Texas where we came in, we didn’t take measurement calibration, we taught the piece of equipment and what happens when you press this button, or how do you accomplish this with this piece of equipment.
It wasn’t measurement calibration training, it was the piece of equipment training to become familiar, to accomplish the procedures that you mentioned.
Weldon:
Exactly. That’s what I was referring to, was training in the calibration equipment itself, not the process of doing the calibration. Back in those days, electronic calibrators were still new in 2000, 2001. There weren’t many people running around with digital equipment for DP calibration.
There were some, let’s just call it better quality digital gauges for pressure, but not anything that we’d call calibration equipment today. That was quite a chore to get some of those techs trained up back in the days when they were used to deadweight testers. Talking about deadweight testers, a lot of them are in use today. They’re still very valid technology.
Can you talk to us about some of the specifics around deadweight testers? Because I think the thought process is it’s a bunch of pieces of iron and tubing, there’s not much you can do to hurt it. Talk to us a little bit about the problems deadweight tester y’all see. What people do to them, what they should be doing, and what they’re not doing?
Brian:
Correct. What’s happening to them again, by and large, is over-pressurization. More so than that, it’s probably debris or sand if there’s any of that that’s not getting filtered out, but it’s more so debris because the deadweight tester is a flow through pressure device. It’s a constant nitrogen source keeping those weights floating.
You have small orifices inside the deadweight tester with the tubing and piping. It does not take much to plug that up. Also in the PK deadweight tester, there’s a ceramic ball that can get a chip knocked out of it, we’ve seen that happen before. That creates a measurement problem because it’s not a smooth surface on that carrier that’s holding the weights.
The storage of the weights is another issue as far as handling them properly, but also storing them properly so that as they’re in transport, they don’t get nicked and dinged.
Weldon:
That’s probably the thing that I’m most familiar with the deadweight testers is, years ago, everybody had a case for those things. It usually was a handmade wooden box in the back, but they were handled with some care.
Just recently in the last year, I was out on a site and we weren’t witnessing calibrations at that time, but the tech went over to grab some tools to loosen up a manifold, and the weights for the deadweight tester were loose at the back of his toolbox. I couldn’t do anything but cringe when I saw that.
Brian: It happens. When you’re setting these trucks up or setting this equipment up, that’s probably one of the most important pieces of equipment to provide storage for is the weights for the deadweight tester above and beyond the deadweight tester.
Weldon: That makes perfect sense. What do you see out there today, Brian? I’ve been away from that, from the field end of the world for, I guess it’s getting close to 10 years, other than doing a few audits here and there. What do you see with folks? How are they setting up their trucks today to properly take care of this equipment?
Brian:
A lot of people are using the three-door pickups, sometimes the four-door pickups. They all leave the third seat in there if it’s a three-door pick up, or they’ll pull the seat out and put some custom cabinetry that’s made specifically to fit in the back of that vehicle.
I’ve also seen people remove that third seat, or even in a four-door pickup, remove all of the back seats, put a panel up along the back wall, and turn that passenger side seat backwards, and work from there inside their pickup, but they have a workstation setup right there.
Some people set it up, they still have the cabinetry for storage of their equipment or holding, but they’ll configure to calibrate out of the driver’s seat and hang their calibration equipment off the steering wheel, have a tray that moves the deadweight tester up close to them, or something that they’ll set it up front.
They’ll leave the back stand on the ground through the third door. They all do a very good job of providing shelves to hold the equipment while they’re doing the testing, but also the storage. The deadweight tester, the case is now come with pockets for the weights.
I’ve seen some people though not use those and build their own pockets inside their drawers, so that way they don’t have to worry about them rattling out while they’re driving down the road or bumping so. It’s a variety.
Weldon:
Excuse me, all choked up here, Brian. Probably choked up over the concept of guys sitting in their truck to do the calibrations. I know that’s extremely popular. I know there’s some companies that endorse it, I know there’s some companies that don’t.
If you’re sitting in your truck doing calibration, that normally means you have a pretty long piece of flexible tubing running from your device under test, back to your truck, to your nitrogen source and your vernier tube or your dead weight tester.
Do you have any thoughts about that one way or another? I know, it’s not strictly in the maintenance of their equipment, but it talks to the topic we’ve got going here.
Brian:
It varies from company to company, what a company will allow. Obviously, the long runs of tubing are not optimal for the pressure change in the stabilization, because once you make a pressure change, even the vernier tube or the nitrogen bottle, you make a step change, you need to let it settle out.
While you’re waiting for that settling time, there could be the hose laying on the ground and it’s cloudy when you started all of that. All of a sudden, the clouds are blown away by the wind or they move away, now the sun comes out and hits the tubing.
Now you’re dealing with the effects of [inaudible 14:36] on the tubing. You can do that, but you need to wait for stabilization. I don’t know how much waiting is going on for stabilization. It’s been a long time since I’ve been out to see that, but once you make those changes, you need to wait for the stabilization.
Weldon:
That’s exactly what I was looking for there, Brian. I’ve seen an example here very recently of a tech using a lot of tubing sitting in his truck, and just immediately make a change, click, make a change, click, make a change, click.
It’s like whoa, hang on a minute here, you’ve got to be watching the data coming back from your device not just what’s your highest or whatever calibrator you’re using. You need to be watching data coming back and make sure both are stabilized before you start hitting that cal button. You’re sure exactly right, heating from the sun, a lot of things can go on there, a lot of things.
What about on the temperature side? What about calibration equipment on the temperature side? What do you see as problems with that? What are folks doing with it?
Brian:
The calibration on the temperature side are typically digital thermometers now. The yellow back mercury thermometers should all be gone by now. There still may be some floating around, but by and large, it’s all the digital thermometers.
The biggest problem that we see are the temperature probes being damaged, and that is typically from being improperly stored, may be thrown in a toolbox, like you said, with the plates from the PK testers in the same drawer as the weights. It’s the storage and care more so than the handling.
They are a pretty robust piece of equipment, so they’re not quite as delicate as the pressure calibration devices. When you improperly store them, that’s where the big issue comes. Dinging up that probe or bending the probe, we’ve seen that even come in as well. That’s the big one, is the storage and the handling for storage.
Weldon:
I see a lot of companies that believe in one piece of test equipment for temperatures, and that’s it. You go out there and a tech will have sometimes even a 12-inch or even an 18-inch temperature probe. You’re out there with that 18-inch temperature probe stuck in the thermal well that’s three inches deep, maybe.
You’ve got a foot hanging out there in the sun or the rolling cold winter, or whatever, and that same concept that we talked about of making sure things stabilize, that sometimes doesn’t happen on temperature also. You get to dip it in, OK, I’ve got a temperature, and that’s it.
Making sure that they’ve got a probe in contact with the bottom of that thermal well, and make sure they don’t have six-foot of that out there exposed to a temperature that’s very different.
Brian:
You’re correct, and that could create some temperature effect problems. Roughly, maybe an inch or two inches of temperature probe sticking out of the top of the thermal well is a good rule of thumb.
Less is always better, but sometimes you have a variety of thermal well differences. You may have to have in some instances maybe more than one or two inches, just to have one piece of equipment.
Unfortunately, we’re not where we can have a variety of pieces of equipment, because we don’t have vehicles large enough to carry it all.
Weldon: What about on the temperature calibration side? Do you see folks these days still using dry wells?
Brian:
We do see some dry wells. We do see a lot of just a reference thermometer that is in the dry well. They’re doing a one point comparison to what is in the pipe.
Some people though will pull the probe that is in the pipe out and put it in a dry well, a temperature source. They call them dry block calibrators, typically. They’ll source a temperature of 32 degrees F or 0 C. Then they’ll run it up to 100 or 120. It depends upon the company, but some companies will go that far.
It’s a function of the company. Do they want to do just a one point verification? Then if the one point verification is good, they’ll leave it.
If there’s a threshold that they set on the one point verification and that threshold is exceeded as far as the differences from the reference standard, the reference thermometer to the unit under test, then they’ll pull it out and do a typically two point calibration, sometimes three point.
Then that’s when they’ll use those dry wells, dry blocks.
Weldon:
I’m a purist with regard to that. I think everything needs at least, at minimum, a two point calibration.
I’ve never come to terms with our industry’s idea that we’re going to stick a reference thermometer in at the same temperature of our process and call it good. That just leaves too much to the imagination when it comes to, is everything linear?
What do we have when we’re not exactly the temperature we need today? Since temperature is such an impact on the volume correction calculation. Well, I won’t get on that soapbox.
With our pressure calibrators, our temperature calibrators, most people think of those, and they think of that as their test equipment. You brought up something when we talked about having this call that always slips my mind.
The calipers and the other equipment used, not only for checking plates, but if you’re going to do tube inspections. Talk to us about what we should be doing with that equipment, how we care for it.
Do we ever verify? Do we ever do any kind of testing on a caliper once it comes out of the factory?
Brian:
The caliper should fall under the same procedures as the pressure calibrator and the temperature calibrator for an annual certification. It’s sent in to be checked and get a certification paper, certification sheet with it.
The storage and handling, again, is one of the big downfalls on the calipers, being thrown in the top drawer of the tool box or thrown in the top drawer of a cabinet. They actually need to be in a form fitting case and stored as such. They’re very a precision piece of equipment.
I’ve spoke to a couple companies that perform calibration services on calipers. They say that they’ve received them back in such shape, where they look like they were being used as a pry bar.
Weldon: Jeez!
Brian: If you need a pry bar, get a screwdriver or a bar. They say they’ve seen them come back bent. Calipers are like a thousand dollars.
Weldon: Wow.
Brian:
They’re actually bent.
There’s that. There’s also if you have any torque wrenches, the same thing. If you’re torquing down any flanges, that can be somewhat critical to make sure that you’ve got consistent torque across all the bolts, especially in higher pressures.
Also, any multimeters or digital electronic equipment. If it’s being used for a measurement variable, then you also need to send those out for calibration as well.
Those are the big ones. The pressure, the temperature, the calipers, and if there’s any electronic verification devices or 4 to 20 milliamps sources or any voltage sources that you’re utilizing to test or to confirm measurement or confirm accuracies, they all need to be sent out.
If you’re not calibrating in-house, send them out to a lab, somebody that can calibrate that equipment.
Weldon:
I like to think, Brian, that when it comes to our transmitters, pressure, temperature, density, DP transmitters, I’d like to think that all of those have moved to the digital communications world, but that’s not the case. There’s still a huge installed base of 4 to 20 equipment out there, probably more often in the facilities of the process plants than anywhere else.
It’s easy to forget that that stuff is there and it’s also easy to, as you mentioned before, treat those loop calibrators as if they didn’t need to be recertified. I don’t know. I think I can count on one hand the number of loop calibrators I’ve seen with the certification sticker on the back.
Brian:
That brings up a good point to lead into the last topic. It’s the certification of the equipment and understanding the certification, and what is a valid certification. If you’re sending it out to a lab, if somebody at your company hasn’t verified the lab that they’re capable or they’re providing qualified calibrations to allow compliance to standard such as API 2011 would be one example.
There’s several other standards that all this falls under, but to make sure that that equipment is being calibrated to satisfy those standards. The individual technician needs to understand the calibration required for each piece of test equipment, how many points? What is a valid measurement? Understanding, is it a correct or valid calibration?
Also qualifying the lab, and then to be able to look and verify the methods and results, more so the results. Probably the as found data and the results being that as found data in the as left data, and going over with a discriminating eye. Looking at that data to make sure it was received in tolerance and it was returned in tolerance.
Especially when it’s received in tolerance with the electronic equipment, it does drift over time. When you receive it back, ideally, you want it to be more than just barely within the tolerance, you want it to be well within the tolerance to compensate for any electronic drift. Even over a year, they can drift. That would be another big one too, is the actual certification and calibration.
Weldon:
Well put, Brian. One thing that I could add to that is, when you receive equipment back from certification or recertification, make sure you review that report. You always hope what you send to the lab is received at the lab still in tolerance, but you need to watch and make sure you don’t have equipment that through damage or just use, drift.
If that equipment went into the shop way out of tolerance, to be recertified, you may need to have a look at how far out of tolerance it was. Is it out of tolerance enough that it potentially impacted any of the equipment you’ve tested recently? Because if you have a transmitter that’s been calibrated with a piece of equipment that was outside of its tolerances, you end up with a mess.
I think that’s a piece that I don’t see happening very often. I don’t see that loop back to say this piece of equipment was bad, do I need to go back and I need to relook? Potentially, even go back and retest some of those meters.
Now, I think when you get a piece of equipment that comes out in the lab says, this piece of equipment is gone, it can’t be brought back into certification, that’s when people stop and think, what happened to the previous devices? Do I need to look at them?
Just that case of where it came into the shop out of tolerance, people don’t always follow up with that and loop back around. The other thing I think that’s important, Brian, and you touched on it there a little bit, I think it might be a topic of a podcast in the future, but understanding the calibration equipment statement of accuracy and the transmitter statement accuracy, and how those two go together.
You need both of those pieces of information if you’re going to do a calibration, this 2011 compliant. That means not hitting the button unless you know you’re out of tolerance. Those calculations either need to be done by the calibration software you do, or the technician needs to do it to literally make that decision, do I hit the calibrate button or do I leave it as is?
Brian: Even before that, a pre-check of the tolerances of the calibration equipment compared to the tolerances of the unit under test to first make sure you’re in compliance with if API 2011 is your driving specification, to make sure you’re in compliance with that before you ever start calibrating.
Weldon: Yep. Great point. What are the words of wisdom do you have for us, Brian, on that equipment? I know you’ve been out there selling it for many years. You’ve been training many years. You’ve been watching people tear it up for many years also. What words of wisdom to the new tech coming into the business about these test equipment?
Brian:
The main thought I’d like to leave everybody with is to take care of your equipment. It’s precision equipment. It’s the storage and handling. I realize there can’t be a case for every piece of equipment, but there can be accommodations made to store the equipment maybe within the cabinets or maybe within cases, even your smaller cases, or combining cases or combining equipment.
The biggest is the storage during transport even before the use, and then the use as well, but it’s more so the storage and the upkeep of it, is the big things that I would like to leave. Also, taking time and waiting for stabilization would be the other one. Above and beyond all the procedures and the calibration steps is just taking time and letting everything normalize and come to equilibrium.
[background music]
Weldon: I think that’s great advice, Brian. Again, I appreciate you being on the podcast with me again. You have a great day, and we’ll see you out in the field somewhere or at the next trade show. Have a great one, Brian.
Brian: Thank you, Weldon.
Weldon:
Thanks for listening, and I hope you enjoyed our episode. If you did, please leave a review at iTunes, Google, or wherever you get your podcast fixes from. A full transcript of this episode along with the contact info on our guests can be found at pipelinepodcastnetwork.com.
If you have suggestions for a topic, reach out to me on LinkedIn or use the contact form at the bottom of each page at pipelinepodcastnetwork.com. Thanks again.
[music]