In this episode of the Oil & Gas Measurement Podcast, host Weldon Wright discusses the important topic of dewpoint measurement with Dr. Sohrab Zarrabian, CEO of ZEGAZ Instruments. Dr. Zarrabian provides insights into both water and hydrocarbon dew point analysis, how dewpoints are measured in the field, and the complexities involved in natural gas systems. The conversation covers the fundamentals of dewpoint, the significance of accurate measurements, and the use of advanced technologies like chilled mirrors for precise dewpoint detection.
Dewpoint Analyzers Show Notes, Links, and Insider Terms
- Dr. Sohrab Zarrabian is the CEO of ZEGAZ Instruments. Connect with Dr. Sohrab on LinkedIn.
- ZEGAZ Instruments designs and manufactures analytical instruments specifically for the oil and gas industry, focusing on hydrocarbon and water dewpoint analyzers for natural gas applications. Their analyzers are used by major pipeline companies, gas refineries, and power generation plants worldwide.
- Dewpoint is the temperature at which liquid droplets first start to condense at a given pressure, as the gas mixture is cooled.
- Hydrocarbon Dewpoint is the temperature that hydrocarbons begin to condense at a given pressure, as a gas mixture is cooled.
- Cricondentherm is the maximum temperature at which both liquids and vapor can coexist for a give gas sample. Worded another way, the temperature above which only gaseous phase hydrocarbons may exist, regardless of pressure.
- Water Dewpoint is the temperature that water begin to condense at a given pressure, as a gas mixture is cooled.
- Bureau of Mines “Chilled Mirror Test”, as defined by ASTM D1142, is a test for determining the water and hydrocarbon dewpoints for gas samples at their flowing pipeline pressure.
- The Chilled Mirror Test uses a small, high pressure chamber with a polished mirror at one end and a viewing window at the other. The sample gas is slowly passed over the mirror at pipeline pressure while the mirror is cooled with a manually regulated refrigerant gas. An operator observes the mirror for the formation of dew, and when it appears, the thermometer reads the dew point temperature. The type of dew that forms can also be determined by an experienced user.
- The test requires a highly skilled operator and takes considerable time to complete.
Dewpoint Analyzers Full Episode Transcript
Weldon Wright: Welcome to episode 36 of the “Oil & Gas Measurement Podcast,” sponsored by GCI, the Gas Certification Institute. For more than 20 years, GCI has been providing measurement fundamentals training and measurement standard operating procedures to the oil and gas industry.
GCI’s instructor-led gas measurement and liquid measurement training classes, coupled with comprehensive measurement standard operating procedures, provide valuable tools for both the back office and field personnel.
[background music]
Announcer: Welcome to the Oil & Gas Measurement Podcast, where measurement professionals, Bubba geeks, and gurus share their knowledge, experience, and likely a tall tale or two on measurement topics for the oil and gas industry.
Now your host, Weldon Wright.
Weldon: Hello. Welcome to the Oil & Gas Measurement Podcast. We are here today with a very, very popular topic, and that is water dew point measurement, hydrocarbon dew point measurement.
We have with us today Dr. Zarrabian of Zegaz, and he is going to talk to us in general about what dew point analysis is, both water and hydrocarbon. He is going to talk to us about how we measure that online and in the field. I think we’re going to find it one of our more popular topics.
Before we start, Sohrab, will you tell us a little bit about yourself, what you do?
Dr. Sohrab Zarrabian: Good afternoon, Weldon. It’s really good to be with you. I’m professionally a physical chemist, chemical physicist, whichever you want to say. I’ve been involved in optical diagnostics for more than 30 years. That’s use of optical technologies to diagnose various things.
I’ve been in the natural gas business for about 20 years. Our company is about 15, 16 years old since the inception in the R&D phase and then bringing product. We’re about 15, 16-year-old company.
I’m the CEO of the company called Zegaz Instruments. We’re located in Frederick, Maryland, about 20 miles outside of Washington, DC.
Weldon: Great. I know that you’re doing a presentation at ASGMT here not too many weeks down the road. As we prepared for this call, I wanted to refresh myself on your products on your website. As I looked at it, I was really impressed by some of the educational material you have there.
One thing I want to make sure we do is that we get your website adress out there and make sure that everyone on the call has it so they can go back and look at it. It’s zegaz, Z-E-G-A-Z, .com.
On their website, they have something that says Zegaz University. There are a couple of great papers on there and there’s a set of videos, including a video from a couple of years back that you did for GPA Midstream on moisture analyzers. I think that’s a great program.
Dr. Zarrabian: That’s it. That’s correct. Our website, zegaz.com, there are two sections of interest for people who want to know more about natural gas measurements in general, but also dew point measurements. One is the Zegaz University tab that has some videos and other technical papers.
Then there’s the blog section. The blog section has some very short articles on various topics of interest. They are very short. They are usually five minutes or less to read, and they go point by point on various topics that people in this industry are interested. I encourage everyone to visit them.
Weldon: That’s a great start. I know — we mentioned this when we prepped the call — when we talk about wellhead to the burner tip, in our industry, that really gets separated into two parts.
It gets separated in the folks on the upstream on the production side, the gathering, the processing, and the treating, the getting the gas ready for delivery, and the folks downstream that are the transportation pipelines, their distribution companies, and then users.
Many times, what those people think of when they hear dew point can be very different depending on what side of that imaginary line they’re on. Do you want to start us off here by talking a little bit about what is dew point?
We hear the term “cricondentherm” thrown around and “water vapor dew point.” There’s hydrocarbon dew it. Set us straight on that before we get started.
Dr. Zarrabian: Great. That’s a great question. Let’s start with just definition of dew point. Dew point is the temperature at which a gas undergoes a phase transition from vapor to liquid. It has units of temperature.
Typically, outside of natural gas industry, when people talk about dew point, they’re talking about moisture dew point. If you’re a fisherman, every morning, the fishing report, it will tell you, “The dew point today is 50 degrees” or 40 degrees or whatever it is.
What that means is that if the temperature of a surface drops below that temperature of, let’s say, 40 degrees, the dew point, moisture will begin condensing. In the natural gas industry, we have two different dew points because not only moisture can condense, but hydrocarbon can condense, too. We have a moisture dew point and then we have a hydrocarbon dew point.
What does dew point depend on, dew point of a gas? It depends on the nature of the gas, what that gas is, and it also depends on its pressure, and to a lesser extent, it depends on what else is in the gas mixture.
In the natural gas industry, we talk about water dew point, which is the fundamental measurement of how much moisture the gas is carrying. When we say the dew point of this gas is, let’s say, minus five, it means that if that gas, at that given condition, pressure, composition, drops below minus five, moisture will start condensing out in the pipeline.
Now, we also have the hydrocarbon dew point, which is the temperature at which the very first hydrocarbons, usually the heaviest ones, will begin to condense out at that given process condition, pressure, composition.
When we say the hydrocarbon dew point, let’s say it’s minus ten, it means that if the pipe hits minus ten, then hydrocarbons will start to liquefy out or to condense out of the pipeline. Obviously, that’s not a good thing. We’re in the gas business. Our pipelines should be gas only, preferably no liquids, because liquids create all sorts of problems.
Weldon: All sorts of problems, but when we get on that upstream section, those liquids, liquid hydrocarbons, are actually the more valuable product, so we’re really worried about them.
One thing is that, when we talk about dew points, whether it’s a water dew point or a hydrocarbon dew point, that dew point is tied to a specific pressure. When we talk about dew point in our air, that’s at one atmosphere. That’s where we live. Right?
Dr. Zarrabian: Absolutely.
Weldon: When we start talking about dew points of the gas in our pipelines, it gets quite a bit more complex, doesn’t it?
Dr. Zarrabian: Yes, it does.
Weldon: One of the things we pretend — and I say “we pretend,” that’s a bad word for it — because we lack other ways of doing it. Many times, we try to calculate a theoretical dew point. We try to calculate a phase curve. Right?
Dr. Zarrabian: Yep.
Weldon: We’ve got to remember that that analysis, in almost all cases, came from a chromatograph, whether it was a chromatograph in a lab or whether it was an online chromatograph, and that chromatograph was operating typically at about 30 psia, about 15 psig, and so the gas in that cylinder can be nothing but a gas at that pressure that goes into the chromatograph.
The relationship between temperature and any kind of liquids, whether it’s water or hydrocarbon liquids in our gas, is that the lower that temperature is and the higher the pressure, the less of it we can have in there. Right?
Dr. Zarrabian: Absolutely.
Weldon: The higher the temperature and the lower pressure, the more of it will hold.
Dr. Zarrabian: The relationship between pressure and dew point is extremely important, both for water and hydrocarbon. Let’s take water, just a simple example. If I take a bit of the air right here, this air has a dew point of about 40 degrees right now here, if I take that air and compress it, liquids will drop out because dew point increases with pressure.
This is what happens in air compressors. You always have to have the drier filter in front. Otherwise, you take the air and you compress it out, liquid will fall out because dew point rises with pressure. That’s for the case of water.
For example, when we have pipelines operating at 800 psi, we can’t really have any more water than maybe 100 ppm, 120 ppm of water, because at that pressure, anything more than that will start to condense out in the liquid.
When you get to a hydrocarbon site, it gets even more complicated because the linear relationship is not just pressure goes up when dew point goes up, there is a regression in the phase diagram, if you look at a typical phase diagram.
Initially, hydrocarbon dew point rises with pressure until it reaches the cricondentherm, and then it regresses with even more increasing pressure. That’s called the retrograde region.
The relationship for hydrocarbon dew point and pressure is a little more complicated. That’s why that term that people mention, cricondentherm, becomes a very important term. That’s the highest dew point at any given pressure. On the phage diagram, it’s really the tip of that belly, the phase diagram belly, it’s the tip of it that’s called the cricondentherm.
Weldon: That is a great explanation. Great.
As we talk about dew points — and we’re going to get to measuring that dew point here in just a few moments — that phase diagram, I teach a few classes and what I find out is a lot of folks in our industry have seen a phase diagram and they have no clue what it means. They also tend to think that that phase diagram is some fixed diagram for all gas.
There’s really not a concept around the fact that that phase diagram can only be calculated if we have a really, really accurate analysis of that gas. Right?
Dr. Zarrabian: Absolutely. You’re right, first of all, the phase diagram changes with the gas composition. You add a little, you subtract a little of one of the components, the phase diagram will change.
More importantly, how accurate is that phase diagram? The phase diagram comes from a calculation using GC-based analysis. If you watch one of those videos on our website or want to read one of the blog sections, there are about five different sources of error when you’re using GC to calculate the phase diagram.
That’s pressure regulation. It’s the inherent error of the GC. It’s the error embedded in the formula, in what they call equation of state formula. It’s also the distribution error. You really need to use a C14 analysis rather than a C6 analysis.
When you look at the phase diagram, even if it corresponds to the gas that you’re interested in, you always have to take it with a grain of salt because there are many sources of error that go in there.
The best phase diagram is one that is experimentally measured, and you can do that with chilled mirrors, and that’s the accurate way to do it. Again, in that paper, we do all those experiments, and there’s lots of useful data in that video as well as the blog section that explains in detail what these errors are.
Weldon: As we see out in the field, we hear people talking about a C6 analysis, a C7, a C8, a C9 analysis, we talk about our chromatographs being set up, whether it’s a chromatograph in our commercial gas lab, or whether it’s that online chromatograph in that plant or that high-volume interconnect site, what a lot of people don’t realize from a commercial term, the value of everything from about pentane zone up is basically the same.
In fact, when the producers of natural gas get paid for their product, there’s usually one pricing for everything that’s pentanes and bigger.
The reason we start talking about these C7, C8, C9 chromatographs, or extended analysis, is because if we’re trying to accurately calculate that phase diagram, if we trust that our chromatograph is as accurate as it could be, we still can be no more accurate than how far out that chromatogram goes.
If we have a C6 chromatograph, we’ve said we really stopped measuring at hexanes and we grouped everything together, and now we’re making an assumption, but that curve of that phase diagram can change dramatically…
Dr. Zarrabian: Absolutely.
Weldon: …if the difference between how much of that was hexanes, how much was decanes, how much was nonanes. That’s why an extended analysis of a properly operated chromatograph is going to get you close. That’s why we talk about these more expensive chromatographs that run a further analysis. That’s a very great point you’ve got us to there.
Dr. Zarrabian: I want to give a shout out to late Professor Kenneth Starling from Oklahoma, who really did a lot of detailed research in this topic and illustrated how these heavy distributions change and can change the phase diagram drastically.
We have a joke here at our company that we said, “If you’re using a C6+ GC to calculate your dew point, save your money. Get a dartboard. Throw a dart.”
[laughter]
Dr. Zarrabian: “You’ll be just as accurate,” which is not accurate at all.
Weldon: Some people are still…
[crosstalk]
Weldon: …better at darts than others.
Dr. Zarrabian: Yeah, absolutely. C6+ GCs, those numbers, they’re not even guesstimates. If you’re relying on that number, be prepared to be in error by 70 to 100 degrees. That’s just not acceptable when you’re operating a natural gas pipeline.
Weldon: This is one of the things that we talk about when we talk about sampling systems. We can’t get information out of our GC without a good sampling system getting to it. Part of trying to develop a good sampling system involves around knowing what we have in the gas and being able to keep that gas warm enough to keep it gas phase as we sample it.
If your very test that you’re trying to run can potentially give you a large variation in that, it’s like, where do you get to? That’s part of the reason that GPA has come out and API has come out and said, “Hey, if you’re basing your sample system heating on an analysis, you need to stay at least 30 degrees above that number.” That’s just trying to be safe, and even that’s not enough. Anyway…
Dr. Zarrabian: Sometimes that’s not enough either. API specifically says, “If you’re trying to measure hydrocarbon dew point,” I forget the number, API 14.1 maybe, but it says, “If you’re trying to measure hydrocarbon dew point,” any measure that’s the preferred method, “measure it with some sort of a chilled mirror device.”
Then it says the next option is use a C14 analysis and an accepted equation of state. Then there are other less useful options. It goes through that. The main option is use a chilled mirror device to measure your hydrocarbon dew point.
Weldon: That’s great. That’s super information. Now that we’ve talked about what dew point is, what cricondentherm is, how do we get it, where are we going to, talk to us a little bit about this legendary chilled mirror dew point test. That’s really the basis for your hardware. Talk to us in general about what that chilled mirror is, and then talk to how you achieve that with your equipment.
Dr. Zarrabian: Very good. First of all, our chilled mirror can measure both hydrocarbon and water dew points. There are two dew point, the chilled mirror measurement for water or there is the reference standard measurement for moisture in anything, not only natural gas, in anything.
If you invent this super-duper moisture sensor and you take it to a standards laboratory like this or somewhere else, what they measure it against is a high-end chilled mirror to see if it’s accurate enough, because it is a reference measurement.
It’s also the reference measurement for hydrocarbon dew points. It is the measurement that validates your equation of state of your formula, software. It is a reference measurement.
Very fortunately, chilled mirror is not a very difficult concept. All chilled mirrors, the ones that were invented 80 years ago to the chilled mirrors that we just invented 10, 15 years ago, all of them have three things in common, and I’ll explain what those are.
The very first thing is that you need to have a surface that you can cool in a controlled manner. Second element is you need to know the temperature of the surface of that mirror or that surface. The third element — this is what really sets our equipment apart from others — is that you need to have a method to detect the onset of condensation.
Let me tell you an experiment that everybody can do at home very easily. Go get a glass of water. Drop a few ice cubes in there. The glass will start to cool down. If you’re watching it very closely, at some point, water beads will appear on the outside of the glass. If you knew the temperature of that glass, you would know the dew point for moisture in your environment.
All of the chilled mirrors have those three things in common. Again, a surface that you can cool in a controlled manner, know the temperature of that surface, and the third is to detect the onset of condensation.
Back 80 years ago when the very first chilled mirrors were being invented — actually, they go a little further back — what they had was a manual chilled mirror. There was a surface in the natural gas industry. You would cool it with some sort of refrigerant like CO2 or propane and a technician would watch it to see when condensation would happen. The method of detection was a human eye.
Obviously, that’s a bit problematic, because not everybody sees the same things and the human eye is not a spectroscopic equipment. It cannot tell the difference between moisture dew point and hydrocarbon dew point, although I know some people who swear they can. [laughs] That was the fundamental problem.
Then there were iterations trying to automate this process. Our equipment, basically, is the latest iteration of automation of that process, where we use spectroscopy to detect the onset of condensation.
Not only we know that condensation happen, but through the use of spectroscopy, we do a quick chemical analysis to know if it was a moisture dew point or if it was a hydrocarbon dew point. That’s really the innovation that our company has brought to the industry.
Weldon: Great. That’s a great summary there. When we talk about this technology, I know we mentioned this in a previous call. You touched around the edges of it. When it comes to measuring hydrocarbon dew point and water dew point, in parts of the natural gas world, we don’t care about both of them, right?
Dr. Zarrabian: Yes.
Weldon: In gathering, we know we have too much water. We’ve got to dehydrate, right?
Dr. Zarrabian: Yeah.
Weldon: By the time we get downstream, we should be dry. You know? We hope…
Dr. Zarrabian: Hopefully.
Weldon: That hydrocarbon dew point is always [indecipherable 22:15] , especially in our powerplants and companies that are really worried about the possibility of introducing liquids. Talk to us a little bit about equipment that can give us both numbers out of the same device. Is that possible?
Dr. Zarrabian: Yes. That is the main thing that our company brought to the table for the industries that, because we’re using spectroscopy to detect the onset of condensation, we can tell if it was moisture dew point or hydrocarbon dew point, and we can measure them simultaneously with the same analyzer.
All we’re doing, conceptually, it’s really simple. Of course, implementation was a little more difficult, but conceptually, it’s very simple. We have an optical crystal, it looks like glass, but we call it an optical crystal, which we cool in a controlled manner using electronic cooling. We have very high-end temperature sensors that we know the temperature of that surface.
Then we’re sending infrared beams on the surface of this mirror, which is probing what is being condensed on it. As soon as condensation happens, it can tell if it was moisture, then it continues on to get the hydrocarbon dew point, or if it was hydrocarbon dew point, then it continues on to get the moisture dew point.
Knowing both is really important, because if you’re running a pipeline, if you’re getting liquid hydrocarbons in your pipeline, that’s a problem. If you’re getting moisture or if you’re getting liquid water, that’s a problem. Both will lead to so many issues.
For example, moisture and hydrocarbon, they always migrate down to the lowest point in the pipeline. They can create blockages. You need to pump harder, because now your effective diameter is smaller. You need to pump harder. There are safety concerns. There’s hydrate formation issues.
If you’re running a turbine to compress this or to generate electricity, creation of these droplets, whether it’s moisture or hydrocarbon, will go in your turbine. It’ll cause multi-million dollar worth of damage.
If you’re a customer of the gas, running a turbine for generation of electricity, it’s really important for you to know what the dew point of the gas is, because once you compress it in the turbine and liquids are created, then it’s too late.
It’s really important all through the process, from the time it’s put into the pipeline, to monitor it, especially if you have a pipeline that has many different on and off ramps, basically, taking in gas, giving it to the customers.
Weldon: As we look at this, we’ve talked a little bit about what that definition of dew point is, what cricondentherm is. We talked a little bit about phase diagrams. We’ve talked a little bit about the gold standard, or really the only standard of the chilled mirror test for measuring dew point. Really, everything else we do is proven back to that standard, right?
Dr. Zarrabian: Yes.
Weldon: As we look at that, we’ve talked through that. Tell us a little bit now, when it comes to putting a online continuous analyzer out there in the field, or if we have a portable device for carrying around, two different things have got to be a concern there.
One of which is how do we get a representative sample into the device? There’s probably some sample conditioning needs to take place. The other is, are there calibration concerns? What needs to happen for long-term maintenance?
Dr. Zarrabian: Really good question. I’ll start with the second part. The calibration concerns. One of the beauties of dew point measurement, it is that it goes to the very first principles of condensation physics.
What that means, dew point is a unit of temperature. As long as your temperature sensor stays in calibration — and those things are dime a dozen now, there are temperature sensors which will keep their calibration more or less forever — the equipment does not need calibration because, think about what you’re doing, you’re cooling the surface, waiting for condensation.
It happens. You look at the thermometer or your digital thermistor and say, “Ah, temperature was 25.” As long as that number 25 comes from a temperature sensor that doesn’t need calibration, then the equipment does not need calibration. That’s the beauty of it, because it goes back to the very fundamentals of physics.
Now, let’s try to contrast that with an aluminum oxide sensor. An aluminum oxide sensor is a electrochemical sensor that changes over time with exposure to natural gas. Moreover, it depends on permeation across the membrane for it to stay constant. Then there is the electronics and everything else. All of those things drift.
Let’s look at the tunable diode laser sensor. Although, it’s a little better for laser sensors, but a lot of those issues still remain, whereas dew point measurement goes to the very first principles of condensation physics. As long as your temperature sensor is a good one that doesn’t drift off — and again, those are plentiful now — you do not need calibration using a dew point sensor.
If you remember those — some of the people may be familiar with the old [indecipherable 28:06] chilled mirror — there was a thermometer. As long as that thermometer was accurate, then…
Weldon: That was it.
Dr. Zarrabian: …that was it. Yep. That’s the beauty of dew point measurement, that it does not need calibration.
Now, regarding how do you handle the sample from the pipeline? I’ll speak about the online analyzers first. Most people, when they want to measure their moisture, the real true figure of merit is the dew point at process pressure.
Let’s say you’re running at 500 psi. What you really care about is, “What is my dew point at 500 psi?” because you need to stay, obviously, above that. Our equipment is designed to take no pressure cut whatsoever to bring the gas in at 500 psi without any sort of pressure cuts.
To measure the dew point analytically at 500 psi with aluminum oxide sensors or laser sensors, all of them operate at atmospheric pressure, you have to take these pressure cuts to come from 500 psi or more down to atmospheric pressure.
You make the measurement, and then there is a calculation that tells you, “Oh, based on this measurement at that atmospheric pressure, your dew point at 500 hundred psi would have been this other thing.” In our devices, there is no such thing. It’s a direct measurement at your process pressure. That’s the figure of merit. That’s what you take to the bank.
Weldon: That’s a key piece there. I don’t know that we have time to dissect it anymore. As we mentioned early on in our discussion, dew point itself, by definition, is dependent upon the pressure you’re at.
Trying to say we’re doing a mathematical correction, if we take a measurement at X pressure and try to mathematically correct it to another pressure, that’s maybe flawed is too strong a word to use for it, but it doesn’t necessarily have a great logic path to it, because we can’t invent more water in our gas. We can get rid of it. We can condense it out. We can have it fall out.
The same with hydrocarbons. We can’t create more hydrocarbons unless they’re being exposed to liquid somewhere in the piping, the sample system, or something like that. There’s a little bit of flaw to that.
I think this has been a really, really great discussion. Anything you want to add before we wrap up?
Dr. Zarrabian: Yeah. One of the things, I just want to recall a conversation I had with one of our customers. Unfortunately, the GC companies of the old, they did a great job selling the C6+ for calculation of hydrocarbon dew point.
I remember we were measuring a customer’s dew point, and it said 20 degrees Fahrenheit. This was somewhere in the Midwest. He was saying that, “No, that’s impossible. My dew point is -80 according to my GC analysis.” I asked him, “Do you have liquids in your pipeline?” He says, “Yes, everybody does.”
I said, “No, not everybody. Only those people who don’t know their dew point accurately, because if you have liquids running into your pipeline, it means that the intake into your pipeline, you didn’t know what its dew point is and at some point it turned into liquid. Now you have slush flowing through the bottom of your pipeline.”
Good analysis, accurate analysis is fundamental to many industries, but particularly to natural gas industry because our industry is really the fundamental industry to power our country and the world.
If you’re not doing things correctly, if you do not have accurate analysis, then everything else goes out the window. You have to strive for accurate analysis. Whether it’s in dew point or all the other parameters that you measure in natural gas industry, analysis is extremely important. Accurate analysis is essential.
Weldon: That’s great. I think it’s a great, great place for us to wrap up here.
Dr. Zarrabian: Great.
Weldon: Dr. Zarrabian, I certainly appreciated having you on the podcast. We’re going to get your information in the show notes with some contact info in that so people can reach out to you. Give us your website one more time so folks can look at the resources on it?
Dr. Zarrabian: Thank you. The website is zegaz.com. That’s Z as in Zebra, Edward, Gary, apple, zebra, .com. The story of the name is something else we can maybe talk about over a beer. That’s the website. Again, I encourage everyone because there’s lots of informative material.
We believe our best customers are important customers, and we put a lot of information on our website, a lot of educational material. I really appreciate your time, Weldon. It’s been good to see you again. By all means, if anybody has any questions, they can reach out to me directly.
[background music]
Weldon: Thanks, Sohrab. I suspect I will see you in about three weeks in Houston at ASGMT.
Dr. Zarrabian: Yes, see you then.
Weldon: Thank you, sir.
Dr. Zarrabian: All the best. Thanks again.
Weldon: I want to thank each and every one of you for listening. I hope you found this episode both interesting and informative. We’ll have a full transcript of this episode and Dr. Zarrabian’s contact info in the show notes on our website, pipelinepodcastnetwork.com.
As always, if you have comments or questions about the episode, suggestions for future topics, or if you’d like to offer yourself up to the podcast mic as a guest, then send me a message on LinkedIn, or shoot me an email at weldon.wright@fpvprime.com.